Introduction:
The UV online monitoring probe can be applied to various occasions that require real-time monitoring of UV power and temperature values. It features a standard RS485 communication interface and MODBUS communication protocol, making it easy to directly communicate with PLC, human-machine interfaces, computers, and other devices for real-time data collection and monitoring.
Features:
1. Extremely wide range, compatible from low light to ultra-strong light, no need to worry about missing data or over-range;
2. Faster response speed, rapid sampling cycle, high accuracy, and strong stability;
3. Dual communication with RS485 and USB, standard MODBUS communication protocol, can meet various secondary development needs;
4. Comes with computer software, no need to purchase a human-machine interface, set up stations on the computer using our software, observe real-time values, which is beneficial for developers to visualize and check if the probe is faulty;
5. Can modify the baud rate and station number, convenient for multi-unit networking;
6. For users who develop their own software but need to monitor UV light intensity, they can directly use our developed computer software for free, which has real-time online monitoring and alarm functions;
7. Can be powered by 5-36V wiring or directly by USB, simple and widely applicable.
Optical Parameters:
Power Range: | 0.1uW/cm²-100,000mW/cm² |
Range Variation: | 1 billion |
Power Error: | ±5% |
Resolution: | 0.1 |
Temperature Measurement Range: | -55℃--125℃ |
Temperature Error: | ±0.5℃ |
Response Speed: | 2048 times/second |
Sampling Cycle: | 10 times/second |
Communication Method: | RS485 / USB dual communication |
Communication Protocol: | MODBUS |
Power Supply Voltage: | Power supply: 5V-36V / USB direct supply: 3.7-4.2V |
Controller Dimensions: | 126.7*83.4*16.7mm (L*W*H) |
Operating Environment: | -20℃-- 125℃, humidity less than 85% no condensation |
Optional Accessories: | Computer monitoring software |
Calibration Cycle: | To ensure measurement accuracy and effectiveness, the recommended calibration cycle is one year |
Probe Selection Table:
Probe Model | E254(UVC) | E365(UVA) | E395(UVA-LED) | E420(Blue Light) |
Wavelength Range | 230-280nm | 320-400nm | 340-420nm | 400-480nm |
Wavelength Peak | 254nm | 365nm | 395nm | 450nm |
Probe Parameters:
Probe Dimensions: | 35*15*14mm |
Sensing Hole Diameter: | φ8mm |
Operating Temperature: | -20℃--150℃ |
Probe Material: | Aluminum |
Standard Cable:
| Standard 1m (customizable for longer lengths, withstands 200℃) |